Abstract
The influence of the charge collection time in silicon detectors on the accuracy of charged particle energy measurements is considered. Calculation and experimental data show that this parameter exerts a significant effect on the selection of the shaping time constants τ of multilayer silicon spectrometers. This is especially important for systems based on Si(Li) detectors, which are designed for detecting long-range charged particles (p, d, t) with energies E ∼ 100 MeV. It is shown that the time constant τ = 1.5 μs ensures a high accuracy in the absolute energy calibration.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.