Abstract

In this study, the distraction length of distraction implant was set as input variable which ranged from 2 to 10 mm. The effect of distraction length on the maximum Von Mises stress in the jaw bones and the implant were evaluated by a finite element method. The results showed that under axial load, the maximum equivalent stresses in cortical bone, cancellous bone, and distraction screw decreased by 5·8%, 8·6%, and 11·0%, respectively, with the changing of distraction length, and under buccolingual load those decreased by 0·3%, 18·0%, and 13·0%, respectively. The data indicate that cancellous bone is more sensitive to distraction length than the cortical bone. Under both loads, the central distraction screw was subjected to the stress concentration and more easily damaged by buccolingual force than by axial force. Distraction implant with distraction length exceeding 8 mm showed relatively better biomechanical behaviour.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.