Abstract

To determine the appropriate tube potential settings for dual-source, dual-energy data acquisition across a range of phantom sizes, and to determine the optimal photon energies for virtual mono-energetic imaging. Water phantoms (15-50-cm wide) containing an iodine test object were scanned on a third-generation dual-source CT scanner using all available tube potential pairs. Virtual mono-energetic images at 40, 50, 60, and 70keV were produced using Mono-energetic Plus. To determine the practical operating parameters for the evaluated CT system, data exclusions were made based on water CT number accuracy, artifacts, and using a noise constraint. Image quality metrics were measured and compared. Excluded tube potential pairs were identified; these were generally at low tube potentials for the low-energy beam and low photon energies. For non-excluded conditions, the highest CNR was obtained using the 70/150Sn setting in phantoms≤35cm at 40keV. 70/150Sn provided optimal iodine CNR below 40cm lateral phantom width at 40keV, while 90/150Sn allowed acceptable image quality in phantoms>40-cm wide at or above 60keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.