Abstract
In order to reduce the number of electrodes in standard 12-lead electrocardiography (ECG) measurement, a number of research works have been conducted. One of the attempts was deriving 12-lead ECG from the EASI lead system, which was initiated by Dower. The previous techniques, including Dower with an improved coefficient, linear regression (LR), and polynomial least square (PLR), calculated the EASI coefficients for one cycle ECG signal, i.e., covering the entire segments. To increase the accuracy, we proposed an approach that calculates the EASI coefficients by segmenting the ECG signal into three segments: PR interval, QRS complex, and ST interval. In other words, coefficients are specific for each segment. Root mean squared error (RMSE) was used to measure the performance. The results showed that our proposed approach outperformed the conventional EASI coefficient calculation in Dower's method and LR, as well as $2^{\text{rd}_{-}}$ , and $3^{\text{rd}}$ -degree PLR.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.