Abstract
A practical implementation of a portable secondary standard for phase modulation (PM) and amplitude modulation (AM) noise at 5, 10, and 100 MHz is described. The accuracy of the standard for both PM and AM noise is +0.14 dB, and the temperature coefficient is less than 0.02 dB/K. The noise floor S/sub phi / (10 kHz) of the standard for PM noise measurements is less than -190 dBC relative to 1 rad/sup 2//Hz at 5, 10, and 100 MHz. The noise floor for AM measurements depends on the configuration. A calibrated level of PM and AM noise of approximately -130+or-0.2 dB relative to 1 rad/sup 2//Hz (for Fourier frequencies from approximately 1 Hz to 10% of the carrier frequency) is used to evaluate the accuracy versus Fourier frequency. Similar PM/AM noise standards are under test at 10 GHz. This new standard can also be used as an alternative to the normal method of calibrating the conversion sensitivity of the PM/AM detector for PM/AM measurements. Some types of time-domain measurement equipment can also be calibrated. >
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.