Abstract

The shunt procedure used for laryngectomized patients undergoing secondary tracheo-esophageal (T-E) puncture is inconvenient and causes stress to the patient. In order to overcome these problems we developed a novel surgical T-E shunt technique using the Groningen voice prosthesis that does not require esophagoscopy or general anesthesia and can be performed in an outpatient clinic. In this procedure, a shunt is created using a pair of nasal forceps with the patient seated. An endoscope with biopsy forceps is used to insert the Groningen voice prosthesis. The procedure is usually completed within 20 min after inducing local anesthesia. Neither the technique itself nor the time taken to complete the procedure differed for T-E and tracheo-neoesophageal (reconstructed with flap) shunting. We believe that this procedure is suitable for patients who are afraid of esophagoscopy and/or are not considered suitable candidates for esophagoscopy and repeated general anesthesia. The procedure is also beneficial for both patients and surgeons with regard to its duration and the cost-effectiveness of treatment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.