Abstract

The goal of the present work was to conduct an initial screening survey of several types of TL and OSL detectors, aimed at searching for the indication of the presence of dose rate effects. The study has been performed on ten different materials: LiF:Mg,Ti; LiF:Mg,Cu,P; CaF2:Dy; Al–P glass; YAP:Mn; CaSO4:Dy; Al2O3:C; BeO; MgB4O7:Ce.Li and quartz. Gamma-ray dose rates ranged from 0.1 mGy/h to 90 Gy/h. No clear evidence of dose rate effects was found for any material. In two cases (MgB4O7:Ce,Li and BeO) some irregularities of the response were observed, which require further investigations but most probably they are not attributable to the dose rate influence.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.