Abstract

In conjunction with the Los Alamos National Laboratory hypervelocity microparticle impact (HMI) team effort to produce higher impact velocities and to understand the physics of crater formation and momentum transfer, we have implemented a low noise microphone as a momentum detector on both the 6 MV Van de Graaff and 85 KV “test stand” particle accelerators. Calculations are presented showing that the impulse response of a circular membrane. When used as a momentum impulse detector, the microphone theoretically may detect impulses as small as 8.8 × 10 −15 N s. Sensitivity of the microphone in this application is limited by the noise threshold of the electronic amplifiers and the ambient microphonic vibration of the system. Calculations lead us to anticipate detection of particles over the full range of the Van de Graaff acceleration capability and up to 7 km/s on the test stand. We present momentum enhancement data in the velocity range between 10 km/s and 20 km/s. Preliminary work is presented on momentum impulse calibration of the microphone using laser-pulse photon momentum as an impulse source.