Abstract
It was the purpose of this finite element analysis to compare the implant-restoration connection created by Procera Implant Bridge (PIB) and conventional screw-retained prostheses with respect to screw loading and gap formation. Finite element models representing a conventional screw-retained restoration and a PIB were set up. A horizontal load of 200N was applied on the restorations while resulting gap formation and screw loading was recorded. Increasing the preload of the retaining screws led to a decrease in gap formation. Smaller gaps were observed in the conventional restorations. Loading of the screws was of comparable magnitudes in both cases. The use of screw-retained implant-supported restorations representing butt-joint connections at the restorative interface may result in increased gap formation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Quintessence international (Berlin, Germany : 1985)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.