This article describes a simple method to determine the presence of metals such as lead and cadmium in plastic toys, paints, and other solids using an atomic absorption spectrometer (AAS) without going through the time-consuming sample preparation steps. Basically, the method involves transferring solid sample to a nichrome wire by placing a small sample on a wire loop or touching the heated wire to the sample and then heating the wire in the AAS flame. By choosing the appropriate radiation source and wavelength, the presence of different metals may be determined qualitatively.