Abstract

An adaptation is presented of method 8261--from the Office of Solid Waste and Emergency Response Test Methods for Evaluating Solid Waste Physical/Chemical Methods (SW-846)-to analyze milk for an expanded list of volatile organic compounds is presented. The milk matrix exhibits a strong affinity for organic compounds and the surrogate based matrix normalization described in method 8261 provided accurate results. This method had the sensitivity necessary to detect volatile organic analytes at or below maximum contaminant levels (MCLs) set by EPA for drinking water. In a survey of milk samples available in Las Vegas, Nevada, 32 of 88 targeted volatile organic compounds (VOCs) were detected. Many of the detected VOCs have not previously been reported and a rationale for their presence in milk is presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.