Abstract

The method of isotropic photorefraction has been used in a trial of refractive screening of 6-9 month old infants. Data are presented on the calibration of the method against retinoscopic measurements and its reliability. In photorefractive screening of 1096 infants under cyclopentolate cycloplegia 5% were found to be hypermetropic (over +3.5 D), 4.5% myopic, and 1.3% anisometropic (over 1 D). These refractive errors were confirmed on retinoscopic follow-up (with the exception of a few anisometropes). Follow-up of controls shows that one small refractive error was missed in 52 infants. We conclude that photorefraction is a valid and practical screening technique. Longitudinal study of infants with refractive errors will assess the value of early detection, in particular for prediction and prevention of strabismus.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.