Abstract
To investigate the image quality and nodules detectability using ultra-low dose (ULD) protocol with iterative model reconstruction (IMR) algorithm when compared to routine low dose (LD) chest CT in lung cancer screening. Chest CT scans were acquired using a 256-slice scanner for 300 subjects. The scan protocol for theULD group was 120kVp/17mAs while forthe LD group was 120kVp/30mAs. All images were reconstructed with filtered back projection (FBP), hybrid iterative reconstruction (HIR) and IMR algorithms. Effective dose was recorded. Image quality assessments were performed by two radiologists. SD of CT attenuation was measured as objective image noise. The number of non-calcified nodules detected in both groups with different reconstruction algorithms were calculated and compared. The effectivedose of ULD group (0.67 ± 0.08 mSv) was about 44% reduced compared with LD group (1.20 ± 0.08 mSv) (p < 0.01). IMR improved image quality and reduced image noise significantly than HIR and FBP in both groups (all, p < 0.01). IMR enabled a higher number of nodule detected compared to FBP and HIR in both LD and ULD groups, especially for solid nodules less than 4 mm. IMR may improve the diagnostic accuracy of ULD CT lung screening with potential nodule detectability improvement. IMR enables significant reduction of the image noise and improvement of image quality in sub-mSv (66% reduction) chest scans.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.