Abstract

A method of testing for bacteriuria with multiple reagent strips for blood, nitrite, and leucocyte esterase was evaluated. Of 669 urine specimens tested, 267 were negative on the reagent strips and so would ordinarily be discarded, creating a reduction in the workload of culturing urine specimens of 40%. Thirteen of these were false negative results, however, which represented 10% of all the samples with definite bacteriuria. Additional time was required for the screening test which reduced potential savings in manpower. Overall, we found costs to be slightly increased. In our laboratory this screening test was not economic and would produce an unacceptable rate of false negative results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.