Abstract

The hypothesis of this study was that acetabular liner vulnerability to scratching from femoral heads, roughened by third bodies embedded in the liner, is not significantly lower for highly cross-linked polyethylene (HXPE) than for conventional polyethylene (CPE). Six CPE and 6 HXPE acetabular liners were each reproducibly embedded with 5 cobalt-chromium-molybdenum (CoCrMo) beads then run for 10 000 cycles in a joint simulator. By visual rank ordering, there was low association between liner scratch severity and polyethylene type. The CPE and HXPE liner scratches were not significantly different in scratch peak-valley height or width or in liner roughness in the vicinity of the embedded beads. This model indicated that high cross-linking of polyethylene does not offer appreciable protection against severe scratching induced by large embedded third-body particles.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.