Abstract

Purpose: To investigate the optimal pulse energy to do sclerostomy with an erbium YAG laser. Materials and Methods: The experiments were performed in enucleated porcine eyes. We changed pulse energy and examined the effects on surrounding tissue. Results: With the increase of pulse energy, the effects of the laser extended to the area surrounding the laser probe. At the threshold energy for doing full-thickness sclerostomy, the total energy was significantly higher than with higher pulse energy. And with pulse energy higher than 2 mJ, the total energy did not show any significant change. Histopathologically, the damaged area around sclerostomy became larger with the increase of pulse energy. Conclusion: The optimal energy to do full-thickness sclerostomy with this system seemed to be 2 mJ.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.