Abstract

We present results obtained with a new scintillation detector of secondary electrons for the variable pressure scanning electron microscope. A detector design is based on the positioning of a single crystal scintillator within a scintillator chamber separated from the specimen chamber by two apertures. This solution enables us to decrease the pressure to several Pa in the scintillator chamber while the pressure in the specimen chamber reaches values of about 1000 Pa (7.5 Torr). Due to decreased pressure, we can apply a potential of the order of several kV to the scintillator, which is necessary for the detection of secondary electrons. Simultaneously, the two apertures at appropriate potentials of the order of several hundreds of volts create an electrostatic lens that allows electrons to pass from the specimen chamber to the scintillator chamber. Results indicate a promising utilization of this detector for a wide range of specimen observations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.