Abstract

With the simultaneous measurement of the temperature signal of a CaWO4 absorber crystal and the scintillation light, it is possible to reject the gamma and electron background in the search for dark matter particles. We report on an alternative new light detector design for the CRESST experiment. For the light detector we use iridium/gold phase transition thermometers with a special circular structure and an Al phonon collector on a silicon substrate. With this detector an energy threshold of 22.3 eV on a 5 σ level was achieved. The light detector was tested in coincidence with a phonon measurement of a 300 g CaWO4 crystal. The phonons from the CaWO4 crystal were measured with an Ir/Au thermometer with a transition temperature of 32 mK.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.