Abstract

A new type lung monitor capable of estimating the distribution of activity in the lungs was developed in order to make correction for the variation in the counting efficiency resulting from the difference of the activity distribution and from the transfer of activity to other organs. The deposition area of activity is roughly estimated by means of image divide counting method. The designed detector consists of a single NaI(Tl)-crystal of 9.5 in. diameter by 0.5 in. thick and seven 3 in. diameter phototubes. The minimum distinguishable amount, defined for localized distribution, is of order of 0.16 nCi 241Am for a 60 min counting period.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.