Abstract
Eighty-seven Nordic Hospital laboratories participated in a joint SCE-NORDKEM follow-up study of the long-term stability of the previously established calibration factors for a number of alpha-amylase routine methods based on six different substrates. Human control materials with 90% pancreatic, 90% salivary, and pure pancreatic alpha-amylases were measured by the participants. The data were plotted before and after calibration of each method using a human pancreatic calibrator with an assigned catalytic concentration of 390 U/l (Phadebas blue starch method, 37 degrees C). As in the previous study, carried out 9 months earlier, the pre-calibration values varied over a six-fold range. The post-calibration values of all methods except those based on a tetraose substrate showed an acceptable inter-laboratory comparability. As a temporary measure, SCE recommends that the Nordic laboratories calibrate the accepted routine methods by their individual calibration factor. Detailed suggestions for calibration procedures and a discussion of the principles of transferability will shortly be published by the SCE in this journal.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scandinavian journal of clinical and laboratory investigation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.