Abstract

Scattering of the evanescent light by a finite-size SiO2 probe is calculated using a point matching method in order to improve the efficiency collecting the near-field optical signal. The scattered-light patterns can be classified into three categories. Category 1 is where the probe tip radius is very small (<λ/13). Category 3 is where the probe radius is very large (≳λ/6). Category 2 is the intermediate case (λ/13<radius<λ/6). When the radius is in category 1, the scattered light is able to be treated as a field radiated by the point dipole induced by the evanescent light at the probe tip. When the tip radius is in category 2 (λ/13<radius<λ/6), in order to obtain the angular distribution of the scattered light, the induced multipoles should be calculated, considering the probe radius and the incident angle. When the tip radius is in category 3 (radius ≳λ/6), the scattered light becomes larger in the parallel direction to a prism surface due to the forward scattering of the evanescent light.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.