Abstract

We measured the scattered radiation received by theatre staff, using high-sensitivity electronic personal dosimeters, during fixation of extracapsular fractures of the neck of the femur by dynamic hip screw. The dose received was correlated with that received by the patient, and the distance from the source of radiation. A scintillation detector and a water-filled model were used to define a map of the dose rate of scattered radiation in a standard operating theatre during surgery. Beyond two metres from the source of radiation, the scattered dose received was consistently low, while within the operating distance that received by staff was significant for both lateral and posteroanterior (PA) projections. The routine use of lead aprons outside the 2 m zone may be unnecessary. Within that zone it is recommended that lead aprons be worn and that thyroid shields are available for the surgeon and nursing assistants.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.