Abstract
Visible emission line coronagraph (VELC) on board ADITYA-L1 mission is an internally occulted mirror coronagraph designed for solar coronal observations over a field of view (FOV) of 1.05 R o to 3 R o . To achieve the proposed science goals, instrument background should be less than 5 ppm over the FOV of VELC. Major contributor toward instrument background is scatter from surface microroughness and particulate contamination over the primary mirror (M1). Hence, a detailed study of scatter through simulations is carried out to arrive at the surface microroughness specifications and surface cleanliness level requirements of M1. Estimation of RMS microroughness correction factor due to finite band width of the profilometer is very important while specifying the RMS microroughness of M1. This paper discusses in detail about scatter simulations, results, and analysis. All simulations are carried out using Advanced System Analysis Program, developed by Breault Research Organization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Astronomical Telescopes, Instruments, and Systems
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.