Abstract
The scatter-reduction properties of a new infinity-focused grid incorporated into a radiographic imaging cassette were analyzed. A polystyrene chest phantom was imaged using the cassette and bedside radiographic procedure. Scatter fractions were measured using a beam-stop technique at several anatomically equivalent locations. The performance of this cassette also was evaluated as a function of the orientation angle. The gridded cassette provided a decrease in scatter fraction from 61% to 49% in the lung and from 87% to 76% in the mediastinum. To obtain equivalent film density when the grid was used, the exposure was increased by a factor of 3.0. While there was a decrease in scatter clean-up as the film cassette was tilted from the perpendicular, the grid performed well out to an angle of 10 degrees. The gridded cassette for portable radiography provides scatter reduction with little sensitivity to alignment. The availability of this device could improve scatter rejection, and therefore, contrast in portable radiographic imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.