Abstract
Local detection of ferromagnetic resonance in magnetically ordered materials is achieved by local thermal modulation of the microwave resonance absorption with a scanning thermal microscope. A scanning thermal microwave resonance microscope was constructed based on a microwave bridge with an X-band cavity and a thermal nano-probe contacted to the sample at the wall of the cavity. The technique has been tested on an epitaxial iron film which exhibited lateral inhomogeneities due to a partial oxidation of iron. Different types of oxidation could be distinguished. The locally resolved ferromagnetic resonance spectra are compared with the result of an integral measurement using conventional detection techniques of the ferromagnetic resonance with the same set-up. The lateral resolution achievable with the thermally modulated ferromagnetic resonance measurements was about 200 nm for the investigated iron film.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.