Abstract

Temperature gradient focusing (TGF) is a recently developed technique for the simultaneous concentration and electrophoretic separation of ionic analytes in microfluidic channels. One drawback to TGF as it has previously been described is the limited peak capacity; only a small number of analyte peaks (approximately 2-3) can be simultaneously focused and separated. In this paper, we report on a variation of the TGF method whereby the bulk flow rate is varied over time so that a large number of analytes can be sequentially focused, moved past a fixed detection point, and flushed to waste. In addition to improved peak capacity, the detection limits of the scanning TGF method can be adjusted on-the-fly, as needed for different samples. Finally, scanning TGF provides a technique by which high-resolution, high-peak-capacity electrophoretic separations can be performed in simple, straight, and short microfluidic channels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.