Abstract

Summary Scanning electron microscopy was used to evaluate changes on the surfaces of polymethylmethacrylate (pmma), silicone, and polyhydroxyethylmethacrylate (hema) intraocular lenses (iol) that had been implanted in the anterior chambers of the right eyes of 15 dogs. Five dogs received pmma iol; 5 dogs received silicone iol; and 5 dogs received hema iol. Twenty-eight days after surgery, the iol were removed and processed for scanning electron microscopy. Nonimplanted iol of each type were processed identically for comparison. Implanted pmma iol had significantly more debris and macrophages on their surfaces than did silicone iol or hema iol. Silicone iol had significantly less fibrin deposition than did pmma or hema iol. Silicone iol had surface defects attributable to handling by surgical instruments. Implanted hema iol had multiple pits that appeared to be caused by biodegradation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.