Abstract

Focused ultrasound is now capable of noninvasively penetrating the intact human skull and delivering energy to specific areas of the brain with millimeter accuracy. The ultrasound energy is supplied in high-intensities to create brain lesions or at low-intensities to produce reversible physiological interventions. Conducting acoustic emission detection (AED) and electroencephalography (EEG) during transcranial focused ultrasound may lead to several new brain treatment and research applications. This study investigates the feasibility of using a novel scalp senor for acquiring concurrent AED and EEG during clinical transcranial ultrasound. A piezoelectric disk is embedded in a plastic cup EEG electrode to form the sensor. The sensor is coupled to the head via an adhesive/conductive gel-dot. Components of the sensor prototype are tested for AED and EEG signal quality in a bench top investigation with a functional ex vivo skull phantom.