Abstract

The Savitzky-Golay(SG) filters are widely used for smoothing and differentiation in many fields namely in elastography, EEG and magnetocardiogram signal processing and mainly in absorption spectroscopy. Despite their exceptional features they were rarely used so far in the ECG signal processing. The aim of this paper is the investigation of their properties in detail from the ECG signal processing aspect. There were considered smoothing and differentiation filters approximating polynomials of zero to fourth order with 3-25-point lengths. The filtering effect and the retaining of small details of signal are opposite requirements and depend on the length and the order of applied polynomial. Taking into consideration 500 Hz sampling rate, size of smallest relevant ECG and requirements of approximating peaks and inflections in ECG signal processing, the best choice of use is the quartic smoothing and differentiation filters with 17-point length

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.