Abstract

We introduce a new concept in radioxenon detection - the radioxenon Array, defined as a system where air sampling and activity measurement is performed at multiple locations, using measurement units that are less sensitive, but on the other hand less costly, and easier to install and operate, compared to current state-of-the-art radioxenon systems. The inter-unit distance in the Array is typically hundreds of kilometres. Using synthetic nuclear explosions together with a parametrized measurement system model, we argue that, when such measurement units are combined into an Array, the aggregated verification performance (detection, location, and characterization) can be high. The concept has been realized by developing a measurement unit named SAUNA QB, and the world's first radioxenon Array is now operating in Sweden. The operational principles and performance of the SAUNA QB and the Array is described, and examples of first measured data are presented, indicating a measurement performance according to expectations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.