Abstract

A saturated solution of PbCrO4 has been proposed as a Calibration Reference Material for Cr(VI). Statistically evaluated results of Cr(VI) determination in a saturated solution of PbCrO4 by electrothermal atomization atomic absorption spectrometry (ETAAS) confirm that the found concentration of Cr(VI), (6.99 ± 0.34 μg l−1) is in complete agreement with the calculated value of Cr(VI) in that solution—7.0 μg l−1. The advantage of such a Calibration Reference Material is that the concentration of Cr(VI) is theoretically predictable on the base of solubility product constant and does not require independent determination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.