Abstract

A new sampling wand concept for ion trap mass spectrometers equipped with discontinuous atmospheric pressure interfaces (DAPI) has been implemented. The ion trap/DAPI combination facilitates the operation of miniature mass spectrometers equipped with ambient ionization sources. However, in the new implementation, instead of transferring ions pneumatically from a distant source, the mass analyzer and DAPI are separated from the main body of the mass spectrometer and installed at the end of a 1.2 m long wand. During ion introduction, ions are captured in the ion trap while the gas in which they are contained passes through the probe and is pumped away. The larger vacuum volume due to the extended wand improves the mass analysis sensitivity. The wand was tested using a modified hand-held ion trap mass spectrometer without additional power or pumping being required. Improved sensitivity was obtained as demonstrated with nano-electrospray ionization (ESI), atmospheric pressure chemical ionization (APCI), and low temperature plasma (LTP) probe analysis of liquid, gaseous, and solid samples, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.