Abstract

Objective To investigate the dose and mechanical properties of medical electron linear accelerators in grass-roots radiotherapy units in Sichuan Province, China via sampling and inspection. Methods A total of eight radiotherapy units in Sichuan Province were selected by sampling, and the tests were performed for the dose and mechanical properties of the medical electron linear accelerators in use. Among these accelerators, there were 5 imported accelerators and 3 domestic accelerators. The test items and methods were determined according to the requirements in GB15213-94. Results Among the 14 test items, the items related to the flatness, symmetry, and overlap of radiation field. The other tests of dose accuracy and mechanical precision achieved good results. Conclusions There is a need to strengthen the daily quality control work for dose and mechanical accuracy of medical electron linear accelerators in grass-roots radiotherapy units in Sichuan Province and perfect the allocation of professional equipment and personnel and training of related personnel. With the support of Sichuan Radiotherapy Quality Control Center, quality control supervision and guidance should cover the whole province. Key words: Accelerator; Radiotherapy dose; Mechanics mechanism; Quality control

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.