Abstract

We report here the design of a sample tube for NMR spectroscopy at medium pressure. The design incorporates a pressure relief mechanism that permits the safe handling of solutions at moderate pressures (up to approximately 10 atm) and has been tested extensively for NMR measurements involving condensed gases as NMR solvents

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.