Abstract

The accurate quantification of the highly unstable intracellular cofactor nicotinamide adenine dinucleotide phosphate in its oxidized and reduced forms demands a thorough evaluation of the analytical workflow and dedicated methods reflecting their solution chemistry as well as the biological importance of their ratio. In this work, we present a workflow for the analysis of intracellular levels of oxidized and reduced nicotinamide adenine dinucleotide phosphate in the yeast Pichia pastoris, including hot aqueous extraction, chromatographic separation in reversed-phase conditions employing a 100% wettable stationary phase, and subsequent tandem mass spectrometric analysis. A thorough evaluation and optimization of the sample preparation procedure resulted in excellent biological repeatabilities (on average <10%, N = 3) without employing an internal standardization approach. As a consequence, the methodology proved to be appropriate for the relative assessment of intracellular levels of oxidized and reduced nicotinamide adenine dinucleotide phosphate in different P. pastoris strains. The ratio of reduced versus oxidized nicotinamide adenine dinucleotide phosphate was significantly higher in an engineered strain overexpressing glucose-6-phosphate dehydrogenase than in the corresponding wildtype strain. Interestingly, a difference was also observed in the nicotinamide adenine dinucleotide phosphate pool size, which was significantly higher in the wildtype than in the modified strain.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call