Abstract
The influence of sample matrix composition on the infrared sensing response to chlorinated hydrocarbons was investigated. The parameters of chemical oxygen demand (COD) and total organic carbon (TOC) were used to characterize water samples from natural and industrial waters. Chlorobenzene was selected as a model target for assessing the effect of the sample matrix, defined by their COD and TOC values, on the infrared optical measurements. It was found that it was possible to directly compare recovery values with TOC and COD values at low concentrations of the latter. However, where TOC and COD values were high, in particular for industrial waters, the influence of sorbing natural organic matter was low compared to the total organic carbon due to the synthetic component of the water. Therefore, when measuring natural environmental waters, it was possible to use COD and TOC values to obtain the true concentration of the analyte in the sample, but for industrial samples, the COD and TOC values were not good indicators.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.