Abstract

An on-line sample enrichment system was designed using monolithic precolumns in microcolumn LC. The monolithic ODS capillary columns were prepared via in situ sol–gel processes. The enrichment efficiency of the monolithic columns was tested by using phthalates as the analytes. The relative standard deviations ( n=6) for the retention time, peak area and peak height were between 0.4 and 1.2%, 0.9 and 5.5% and 0.4 and 3.9%, respectively. The system was linear ( R 2>0.99) within the working sample concentration and sample volume ranges. Comparing to 0.2 μl injection with a typical sample injector, the theoretical plate number of a same separation column was increased by 3–6-fold when the precolumn unit was used for sample injection. The recoveries of the analytes were between 88 and 120%, and the sample volume that could be injected into the system was increased up to 5000-fold. The limits of detection were improved by more than 2000-fold and were between 0.21 and 0.87 ng ml −1 even with a UV absorbance detector. This system was applied to the determination of phthalates contained in laboratory distilled water and tap water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.