Abstract

The use of modulated temperature differential scanning calorimetry (TMDSC) and differential scanning calorimetry (DSC) in the measurement of the glass transition temperature (Tg) in polymer-water systems presents several important problems. These include the presence of water evaporation endotherms, partial water evaporation during scanning, changes in pan integrity due to vapour pressure developing in the pan headspace during analysis, and possible interaction between water and polymer at high temperatures. As a result, in most of the cases, only apparent Tg values can be obtained. In this study, TMDSC and DSC were used to determinate the thermal behaviour of methylmethacrylate copolymer-water systems. The samples were previously equilibrated at different relative humidities (RH) from 0 to 97% RH. Three different pan arrangements were used. In addition, thermogravimetric analysis (TG) was carried out to determine the initial amount of water in the sample. None of the pan arrangements was entirely suitable for the study of these systems. When sealed pans were used, the plastification effect of water was observed. Some evidence of degradation was also observed in which water and methylmethacrylate appeared to play roles.