Abstract

BackgroundThe aims of this study were to establish a normal database (NDB) for semiquantification of dopamine transporter (DAT) single-photon emission computed tomography (SPECT) with [123I]FP-CIT on a cadmium zinc telluride (CZT) camera, test the preexisting NaI-derived NDB for use in CZT scans, and compare the diagnostic findings in subjects imaged with a CZT scanner with either the preexisting NaI-based NDB or our newly defined CZT NDB.MethodsThe sample comprised 73 subjects with clinically uncertain parkinsonian syndrome (PS) who prospectively underwent [123I]FP-CIT SPECT on a CZT camera according to standard guidelines with identical acquisition and reconstruction protocols (DaTQUANT). Two experienced readers visually assessed the images and binarized the subjects into “non-neurodegenerative PS” and “neurodegenerative PS”. Twenty-five subjects from the “non-neurodegenerative PS” subgroup were randomly selected to establish a CZT NDB. The remaining 48 subjects were defined as “test group”. DaTQUANT was used to determine the specific binding ratio (SBR). For the test group, SBR values were transformed to z-scores for the putamen utilizing both the CZT NDB and the manufacturer-provided NaI-based NDB (GE NDB). A predefined fixed cut-off of -2 was used for dichotomization of z-scores to classify neurodegenerative and non-neurodegenerative PS. Performance of semiquantification using the two NDB to identify subjects with neurodegenerative PS was assessed in comparison with the visual rating. Furthermore, a randomized head-to-head comparison of both detector systems was performed semiquantitatively in a subset of 32 out of all 73 subjects.ResultsCompared to the visual rating as reference, semiquantification based on the dedicated CZT NDB led to fewer discordant ratings than the GE NDB in CZT scans (3 vs. 8 out of 48 subjects). This can be attributed to the putaminal z-scores being consistently higher with the GE NDB on a CZT camera (median absolute difference of 1.68), suggesting an optimal cut-off of -0.5 for the GE NDB instead of -2.0. Average binding ratios and z-scores were significantly lower in CZT compared to NaI data.ConclusionsUse of a dedicated, CZT-derived NDB is recommended in [123I]FP-CIT SPECT with a CZT camera since it improves agreement between semiquantification and visual assessment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.