Abstract

An enzymatic biosensor was developed for salicylic acid (salicylate ion) determined using a Clark type gas diffusion electrode and two enzymes (tyrosinase and salicylate hydroxylase) entrapped in a cellulose triacetate membrane. After optimization, the method was applied to the determination of salicylic acid in cow urine. Relatively good recoveries were achieved, between about 83% and 109%, using the calibration curve, and acceptable precision (R.S.D. about 8%). The method is now being tested for the determination of salicylic acid contained in commercially available drug specialities or galenic products. So far agreement with nominal values has been found to be between 75% and 110% with a R.S.D. of less than 8%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.