Abstract

A Si(111) winged crystal has been designed to minimize anticlastic bending and improve sagittal focusing efficiency. The crystal was thin with wide stiffening wings. The length-to-width ratio of the crystal was optimized by finite element analysis, and the optimal value was larger than the `golden value'. The analysis showed that the slope error owing to anticlastic bending is less than the Darwin width. The X-rays were focused two-dimensionally using the crystal and a tangentially bent mirror. The observed profiles of the focal spot agreed well with the results of a ray-tracing calculation in the energy range from 8 to 17.5 keV. X-ray diffraction measurements with a high signal-to-noise ratio using this focusing system were demonstrated for a small protein crystal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.