Abstract
YAG sclerostomy was performed in 15 eyes of nine cynomolgus monkeys. Individual pulse energies ranged between 10 and 135 millijoules. Total pulse energy ranged between 2,000 and 36,000 millijoules in order to test the extreme range of energy that might be required. Both acute effects and chronic effects were studied in the corneal endothelium, the lens capsule, the iris, the ciliary body, the retina, and the sclera. Techniques included flat preparations of the corneal endothelium, paraffin sections for light microscopy, and scanning and transmission electron microscopy. Our conclusion is that individual pulse energies of up to 135 millijoules with a total of up to 36 joules of energy are safe in monkey eyes when performing YAG sclerostomy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.