A statistical analysis of ceramic implant failure is presented. It is shown that the majority of failures occure within the first 12 months after implantation. From this characteristic distribution the prediction of the total failure rate is possible even for the latest years of production. There have been considerable improvements of the reliability in the past 5 years. Furthermore, there is evidence that the failure rate strongly depends on ball head diameter. For the largest ceramic ball head of 36mm diameter up to now no failure has been reported. From these results it is concluded that ceramic implant failure is in general released by individual circumstances. In particular, the probability of failure may be strongly increased in the case of limited mobility and an increased risk of impingement. In the case of a well developed specific surgeon training and right choice of the ceramic implant system the ceramic components provide an extremely reliable and safe solution for the patient.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.