Abstract

Nasal Intermittent Positive Pressure Ventilation (NIPPV) is an effective therapy for infants in respiratory distress. We here report the safety of a novel, low-cost, non-electric bubble NIPPV device in comparison with bubble NCPAP. At Paramitha Children's Hospital (Hyderabad, India), preterm (n = 60) neonates with moderate respiratory distress were pragmatically allocated to bubble NCPAP (5-8 cm H2O) or bubble NIPPV (Phigh 8-12 cm H2O/Plow 5-8 cm H2O) based on staff and equipment availability. Primary outcomes to assess safety included clinically relevant pneumothorax, nasal septal necrosis, or abdominal distention. One patient in each arm developed minor nasal septal injury (grade 3 on NCPAP, grade 2 on NIPPV); no patients in either arm developed a clinically significant pneumothorax or abdominal distention. The similar rates of nasal septal injury, pneumothorax and abdominal distention suggest that bubble NIPPV has a similar safety profile as bubble NCPAP for preterm infants in respiratory distress.