Abstract

To evaluate the safety of 3 Tesla (T) magnetic resonance imaging (MRI) in patients with auditory brainstem implants (ABI) with the magnet removed at implantation and report incidence of complications. Retrospective chart review. Tertiary neurotology ambulatory practice. Patients with diagnosis of Neurofibromatosis, type 2 (NF2) with functional ABIs. Observational recordings. Of the 89 patients meeting inclusion criteria, 7 patients underwent 3T MRI, with a total of 39 scans done. Three patients had 1 scan each, one patient had 4 scans, one patient had 5 scans, one patient had 6 scans, and one patient had 21 scans. The mean time between ABI placement and first 3 T scan was 118 ± 73 months. The most common indication for imaging was surveillance of NF2 lesions. The most frequent scans were MRI brain (25.6%), followed by MRI of cervical (15%), thoracic (15%) and lumbar (15%) spine, and MRI IAC (8%). There were no reported complications for any of the scans. No scans were interrupted due to patient discomfort. There were no device malfunctions. 3 T MRIs are safe in patients with ABIs as long as the magnet is removed. It is recommended that the magnet be removed at the time of implantation in all NF2 patients, who require frequent surveillance.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call