Abstract
Major safety deficiencies of older mobile hypothermia units are described. Problems include absence of high or low temperature output limitation or of visual warning of abnormal operating temperatures. Modifications described provide two high temperature thermostats and one low temperature thermostat, which shut the unit down if the temperature limits are exceeded. Visual and auditory alarms were also added. Operators of mobile thermal units should be aware of the major safety deficiencies of some older models, and should insist on either modification, where appropriate, or replacement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.