Abstract

The goal of this study is to assess the safety and efficacy of mechanical lead extraction utilizing the Evolution system. Compared with other techniques commonly used for lead extraction, data regarding the safety and efficacy of mechanical lead extraction using the Evolution system is limited and needs further evaluation. Between June 1, 2009 and September 30, 2016, we retrospectively analyzed 400 consecutive patients who exclusively underwent mechanical lead extraction utilizing the Evolution system. A total of 400 patients underwent mechanical lead extraction of 683 leads. Mean age of extracted leads was 6.77 ± 4.42 years (range 1 to 31 years). The extracted device system was an implantable cardioverter-defibrillator in 274 patients (68.5%) and a pacemaker system in 126 patients (31.5%). Complete lead removal rate was 97% with a clinical success rate of 99.75%. Incomplete lead removal with<4-cm remnant was associated with older leads (lead age >8years). Failure to achieve clinical success was noted in 1 patient (0.25%). Cardiac papillary avulsion, system-related infection, and cardiac tamponade were the major complications noted in 6 patients (1.5%). Minor complications were encountered in 24 patients (6%), of which hematoma requiring evacuation was the most common minor complication. There were no patient deaths. In our single-center study, lead extractions utilizing the Evolution mechanical lead extraction systemwere safe and effective and resulted in high clinical and procedural success, with low complication rates andnofatalities.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call