Abstract

Background. Systemic protamine sulfate for heparin reversal after cardiopulmonary bypass (CPB) is associated with uncommon, but life-threatening adverse reactions. Methods. In a prospective randomized 3-day outcomes study, a heparin removal device (HRD) group (n = 12; 60-, 80-, 100-kg subgroups) was compared with a matched systemic Protamine group (Protamine; n = 6) for safety and efficacy using an adult swine model of CPB (60 minutes, 28°C). Results. HRD run time was 25 to 38 minutes depending on weight without complications. After HRD, heparin concentration decreased from 4.77 ± 0.17 to 0.45 ± 0.06 U/mL (activated clotting time [ACT] 776 ± 83 to 180 ± 12 seconds), and in Protamine, 3.94 ± 0.63 to 0.13 ± 0.02 U/mL (ACT 694 ± 132 to 101 ± 5 seconds) (p = 0.01 between groups, but no significant differences 60 minutes later). No significant difference between HRD and Protamine to 72 hours was seen in plasma-free hemoglobin C3a, heparin concentration, thromboelastogram index, platelet count, activated partial thromboplastin time, anti-thrombin III, fibrinogen, ACT, and tissue histology. Conclusions. In a prospective randomized outcomes study, HRD achieved predictable reversal of systemic heparinization after CPB with no difference in safety or outcomes compared with protamine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call