Abstract

Pre-implant ECG screening is performed to ensure that S-ICD recipients have at least one suitable sensing vector, yet cardiac over-sensing remains the commonest cause of inappropriate shock therapy in the S-ICD population. One explanation would be the presence of dynamic variations in ECG morphology that result in variations in vector eligibility. Adult ICD patients had a 24-h ambulatory ECG performed using a digital Holter positioned to record all three S-ICD vectors. Using an S-ICD simulator, automated screening was then performed at one-minute intervals. In vectors with a mean vector score>100 (the accepted value for a passing vector when screened on a single occasion), the percentage of all screening assessments that passed, eligible vector time (EVT), was calculated. EVT was compared statistically to QRS duration, corrected time to peak T (pTc) and mean vector score. Ambulatory monitoring was performed in 14 patients (mean age 63.7 ± 5.2 years, 71.4% male) with 42 vectors analysed. In 19 vectors the mean vector score was>100. Within this "passing" cohort EVT varied between 42.7% and 100%. In 7/19 (37%) the EVT was<75%. A negative correlation was found between QRS duration and EVT (Pearson correlation -.60, p=.007). No correlation was found between EVT and mean vector score or pTc. Vector eligibility is dynamic. When "passing" vectors are subjected to repeated screening, 37% are found to be ineligible, more than a quarter of the time. Further investigation is required to determine the clinical significance of these findings.