Abstract

The ZEUS calorimeter has now been operated for almost two years, collecting ep collision data at HERA. This uranium-scintillator calorimeter covers 99.8% of the solid angle and has an energy resolution of 35%//spl radic/E for hadrons and 17.5%//spl radic/E for electrons. It is intercalibrated at the level of 1% with the natural uranium radioactivity. It achieves better than 1 ns timing resolution for energy deposits greater than 4.5 GeV. It has an excellent noise performance with no coherent noise and a dynamic range equivalent to 16 bit. With the readout of its /spl sim/12000 channels, operated in pipelined fashion running at 10 MHZ for 5 /spl mu/s, and the trigger, it is a major piece of hardware which has to be kept operational and well calibrated. We review the detector as well as its readout and discuss the methods used for calibration and monitoring. The stability, the energy as well as the time calibration will be covered. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.